Ultrasound
Ultrasound (above kHz; medical uses – MHz) is generated and detected by a piezoelectric crystal, which vibrates when an alternating p.d. is applied and produces an e.m.f. when squeezed by returning echoes. Pulses reflect at boundaries between tissues. The fraction reflected depends on the acoustic impedances : . A big mismatch reflects almost everything, so gel is used to couple the transducer to the skin. Ultrasound is attenuated exponentially: .
On the Edexcel 9PH0 syllabus (First teaching 2015): 2 points this lesson covers
- 88 Reflection and transmission at boundaries
- 89 Pulse-echo technique
Summarised in our own words from the official syllabus. Check the exact wording in the board's specification.
Learning objectives
- Explain how a piezoelectric transducer produces and detects ultrasound.
- Use and the intensity reflection coefficient.
- Explain impedance matching and the use of coupling gel.
- Use and find depths from echo times.
Before an ultrasound scan, the doctor spreads cold gel on your skin. Why can't they just press the probe on directly?
Core idea
Ultrasound scans build pictures from echoes. Sound bounces off each boundary between tissues, and timing the echoes tells how deep each boundary is. How strongly it bounces depends on how different the two materials are.
. Reflected fraction . Depth .
1. Producing ultrasound
- Piezoelectric crystal: vibrates with an alternating p.d.
- Produces a p.d. when echoes hit it: one transducer sends and receives.
2. Reflection
- where
- acoustic impedance (kg m⁻² s⁻¹).
, .
- Goal
- Reflected fraction.
- Strategy
- Formula.
- Powers of 10 cancel.
- Square.
3. Depth from echoes
Echo after 65 μs; m s.
- Goal
- Depth.
- Strategy
- Halve the round trip.
- There and back.
- Halve.
Why is a coupling gel used in ultrasound scans?
Summary
- ; reflection formula.
- Depth .
- Acoustic impedance
- Density × speed of sound.
Also on your syllabus
- Fibre optics and endoscopy: light travels along a thin glass fibre by total internal reflection at the core–cladding boundary (the cladding has a lower refractive index). An endoscope has a coherent bundle (fibres keep the same relative positions at both ends) to carry the image back, and a non-coherent bundle to carry light in. Uses include viewing inside the body and keyhole surgery.
- MR scanner: the patient lies in a strong uniform magnetic field, which lines up the spins of hydrogen nuclei (protons). A radio-frequency pulse at the precession (Larmor) frequency flips them; as they relax back they emit radio signals, picked up by coils. Gradient fields make the frequency depend on position, so a computer builds a cross-section. It uses no ionising radiation and gives excellent soft-tissue detail, but it is expensive, slow, noisy and cannot be used with some metal implants.